Intermodule Measurements

To adjust for minimum skew between two modules involved in an intermodule measurement

8Record the differences shown by the two modules.

You can use the X and O markers to measure the differences in delays.

9Select the Intermodule Menu.

10Select Skew, then enter a skew correction value for one of the modules using the knob or the keyboard.

11Return to the module waveform display and recheck the skew adjustment.

You will need to repeat steps 7 through 11 as needed until the trigger events on the two waveforms match as closely as possible.

Before making an intermodule measurement, you should remove skew between the modules to ensure that a simultaneous arm signal to both modules results in captures of the same events around the trigger.

Skew Removal Between Timing Analyzer and Oscilloscope

2–28