HP 16500C, 16501A LOGIC manual Verifying Setup Time for Memory Address

Models: 16500C 16501A LOGIC

1 130
Download 130 pages 6.88 Kb
Page 25
Image 25

Triggering

To verify that the chip select line of a memory chip is strobed after the address is stable

To verify that the chip select line of a memory chip is strobed after the address is stable

1Select the timing analyzer Trigger menu.

2Define a term called ADDRESS to represent the address in question and the Edge1 term to represent the asserting transition on the chip select line.

You can rename the Edge1 term to suit the problem, for example, to MEM_SEL.

3Under Timing Sequence Levels, enter the following sequence specification:

Find “ADDRESS” > 80 ns

TRIGGER on “MEM_SEL” 1 time Else on “ADDRESS” go to level 1

Verifying Setup Time for Memory Address

1–15

Page 25
Image 25
HP 16500C, 16501A LOGIC manual Verifying Setup Time for Memory Address