Intermodule Measurements
To see the status of a module within an intermodule measurement
To see the status of a module within an intermodule measurement
1Select the Intermodule menu.
2Find the name of the module under the “Modules” list, and read the status under the module name.
The status can be either Running or Stopped. You can interpret these indications as follows:
∙If a module was running and is now stopped, assume it received its arming signal, triggered, and finished its measurement properly.
∙If a module located below a stopped module on the intermodule configuration tree has received an arming signal and is still running, either it is still waiting to satisfy its trigger specification or it has not captured enough information to fill its memory.
∙If a module below a running module on the intermodule configuration tree has not received its arming signal, it will not begin running until the upper module finds its trigger condition.