HP E1368A manual ∙ Number of Scans Use only numeric values between 1

Models: E1368A

1 86
Download 86 pages 17.67 Kb
Page 51
Image 51

ARM

The ARM subsystem selects the number of scanning cycles (1 through 32767) for each INITiate command.

ARM

:COUNt <number> MIN MAX

:COUNt? [MIN MAX]

:COUNt ARM:COUNt <number> MIN MAX allows scanning cycles to occur a multiple of times (1 to 32767) with one INITiate command when INITiate:CONTinuous OFF 0 is set. MIN sets 1 cycle and MAX sets 32,767 cycles.

Parameters

 

Parameter Name

Parameter Type

Range of Values

Default Value

 

 

 

 

 

 

number

numeric

1 - 32767 MIN MAX

1

 

 

 

 

 

 

 

Comments

Number of Scans: Use only numeric values between 1 and 32767

 

for the number of scanning cycles (default is 1).

 

 

Related Commands: ABORt, INITiate:IMMediate

 

 

*RST Condition: ARM:COUNt 1

 

Example Setting Ten Scanning Cycles

This example sets a Microwave Switch for 10 scans of channels 00 through

02.When the scan sequence completes, channels 00 through 02 (switches

00through 02) are closed.

ARM:COUN 10

Set 10 scans per INIT command.

SCAN (@100:102)

Scan channels 00-02.

INIT

Start scan, close channel 00.

Chapter 4

HP E1368A/69A/70A Command Reference 51

Page 51
Image 51
HP E1368A manual ∙ Number of Scans Use only numeric values between 1, For the number of scanning cycles default is