ARM
The ARM subsystem selects the number of scanning cycles (1 through 32767) for each INITiate command.
ARM
:COUNt <number> MIN MAX
:COUNt? [MIN MAX]
:COUNt ARM:COUNt <number> MIN MAX allows scanning cycles to occur a multiple of times (1 to 32767) with one INITiate command when INITiate:CONTinuous OFF 0 is set. MIN sets 1 cycle and MAX sets 32,767 cycles.
Parameters
| Parameter Name | Parameter Type | Range of Values | Default Value |
|
|
|
|
|
| number | numeric | 1 - 32767 MIN MAX | 1 |
|
|
|
|
|
|
| |||
Comments | ∙ Number of Scans: Use only numeric values between 1 and 32767 | |||
| for the number of scanning cycles (default is 1). |
| ||
| ∙ Related Commands: ABORt, INITiate:IMMediate |
| ||
| ∙ *RST Condition: ARM:COUNt 1 |
|
Example Setting Ten Scanning Cycles
This example sets a Microwave Switch for 10 scans of channels 00 through
02.When the scan sequence completes, channels 00 through 02 (switches
00through 02) are closed.
ARM:COUN 10 | Set 10 scans per INIT command. |
SCAN (@100:102) | Scan channels |
INIT | Start scan, close channel 00. |
Chapter 4 | HP E1368A/69A/70A Command Reference 51 |