:OPERation:ENABle STATus:OPERation:ENABle <number> enables the Standard Operation Status Register to set a bit in the Status Byte Register. For the Microwave Switch, when bit 8 in the Standard Operation Status Register is set to 1, bit 7 in the Status Byte Register is set to 1.

Parameters

Parameter Name

Parameter Type

Range of Values

Default Value

 

 

 

 

number

numeric

0 - 32767

0

 

 

 

 

 

 

 

 

Comments Summary Bit: If bit 8 is 1 in the Standard Operation Status Enable Register, a transition of Event Register bit 8 to 1 will cause the associated summary bit to be true.

Maximum Value Entered: The maximum decimal weighted value used in the Microwave Switch module is 256 (bit 8 set true).

Setting Bit 7 of the Status Byte Register:

STATus:OPERation:ENABle 256 sets bit 7 of the Status Byte Register to 1 after bit 8 of the Standard Operation Status Enable Register is set to 1.

Related Commands: [ROUTe:]SCAN

Example Enable the Status Byte Register Bit 7

STAT:OPER:ENAB 256

Enables bit 7 of the Standard

 

Operation Status Enable Register,

 

which enables bit 7 of Status Byte

 

Register.

:OPERation:ENABle? STATus:OPERation:ENABle? returns the bit value of the Standard

Operation Status Enable Register.

Comments Output Format: Returns a decimal weighted value from 0 to 65535 indicating which bits are set true.

Maximum Value Returned: The maximum decimal weighted value used in the Microwave Switch module is 256 (bit 8 set true).

Example Query the Standard Operation Status Enable Register

STAT:OPER:ENAB?

Queries the Standard Operation

 

Status Enable Register.

Chapter 4

HP E1368A/69A/70A Command Reference 63

Page 63