Example: Multiple
Channel Switching
using EXTernal
POWER
This example illustrates closing and opening coaxial switch channels 00 through 04 in a HP E1369A Microwave Switch Driver module. For the example, the switches are not installed on the HP E1369A card, and have 24Vdc drive requirements (HP 33311B or 8762B). Figure
To close channels 00 through 04, execute:
CLOS (@100:104) | Connect channels 00, 01, 02, 03, |
| and 04 switch port 2 to port C, and |
| terminate port 1 into 50; 1 is the |
| card number and |
| channel numbers. |
To open channel 00 through 04, execute: |
|
OPEN (@100:104) | Connect channels 00, 01, 02, 03, |
| and 04 switch port 1 to port C, and |
| terminate port 2 into 50; 1 is the |
| card number and |
| channel numbers. |
Figure 3-4. Example: Multiple External Channel Switching using External Power