OUTPut
The OUTPut subsystem enables or disables the "Trig Out" port of the HP
E1300/E1301 Mainframe or on the HP Command Module.
Subsystem Syntax | OUTPut |
|
|
|
| [:STATe] <state> |
|
| |
| [:STATe]? |
|
|
|
[:STATe] | OUTPut[:STATe] <state> enables/disables the "Trig Out" port on the rear | |||
| panel of the HP E1300A/E1301A Mainframe or on the HP Command | |||
| Module. OUTPut[:STATe] ON 1 enables the port and OUTPut[:STATe] | |||
| OFF 0 disables the port. |
|
| |
Parameters |
|
|
|
|
|
|
|
|
|
| Parameter Name | Parameter Type | Range of Values | Default Value |
|
|
|
|
|
| state | boolean | 0 1 ON OFF | OFF 0 |
|
|
|
|
|
|
| |||
Comments | ∙ Enabling "Trig Out" Port: When enabled, a pulse is output from | |||
| the "Trig Out" port after each switchbox channel is closed. If |
disabled, a pulse is not output from the "Trig Out" port after channel closures. The output pulse is a +5V
∙"Trig Out" Port Shared by Switchboxes: When enabled, the "Trig Out" port may be pulsed by any switchbox each time a scanned channel is closed. To disable the output for a specific switchbox, send the OUTPut[:STATe] OFF or OUTPut[:STATe] 0 command for that switchbox.
∙Related Commands: [ROUTE:]SCAN, TRIGger:SOURce
∙*RST Condition: OUTPut[:STATe] OFF (port disabled)
Example Enabling "Trig Out" Port
OUTP:STAT ON | Enables "Trig Out" port to output |
| pulse after each scanned channel is |
| closed. |
[:STATe]? OUTPut[:STATe]? queries the present state of the "Trig Out" port. The command returns 1 if the port is enabled, or 0 if disabled.
Example Query "Trig Out" Port Enable State
This example enables the "Trig Out" port and queries the enable state. The
OUTPut[:STATe]? command returns 1 since the port is enabled.
OUTP:STAT ON | Enables "Trig Out" port for pulse |
| output. |
OUTP:STAT? | Query port enable state. |
Chapter 4 | HP E1368A/69A/70A Command Reference 57 |