Using the System
October 2002 2-3
Calibrating the Printer
You need to calibrate the printer when you start the system up each day. You also
need to calibrate the printer when
you change paper
print quality is questionable
the temperature at the site changes more than 5° F (2.8° C)
if running more than one shift, at the beginning of each shift
If you are calibrating at any time other than during the daily startup procedure:
Before you begin, make sure the printer is not receiving printing co mmands from
the host.
NOTE: Most windows displayed in this section are from the Windows NT version
of the software. The windows for the Macintosh version of the software are
similar.
For information about editing the calibration settings (for example, y ou may want
to display a calibration graph only if calibration is out of toleranc e), see “Editing
the Calibration Configuration” on page Appendix C:-19.
To calibrate the printer:
1. Check that:
the densitometer is connected and calibrated
the printer status is “Online and Ready”
the processor is in control
the calibration application is running
2. Select the icon for the LED Prin ter on the KODAK Device Calibration screen
and click the Calibrate icon. The calibration software needs complete control
of the printer to successfully calibrate the printer.
IMPORTANT:If no icon appears on the KODAK Device Calibration Screen, you
will need to add a calibration device. See “Adding a Device” on
page C-6.
Calibrate
icon