System Operating Procedures
Entering the Flow Rate
1.Measure the flow rate. See “Measuring the Flow Rate” above.
2.Press Set twice to enter the Calibrate mode.
3.Use the Up or Down Arrow to enter the measured flow rate.
4.Record and save the calibration values so they can be reentered if the memory is cleared.
5.Press Set once or press a different blue function key to return the screen to View mode.
Managing the Development Time
IMPORTANT: To prevent paper jams or image artifacts, operate the processor development time within the recommended
Viewing the Development Time
Press Drive to access the Drive Motor screen.
NOTE: If the drive is running, the current development time appears on the bottom line. It is displayed in minutes and seconds. If not, the screen shows a development time of
Setting the Development Time
1.Select the Drive Motor screen.
2.Press Set to access the Drive screen in Set mode.
3.Use the Up or Down Arrow to change the set development time.
4.Press Set three times or press a different blue function key to return the screen to View mode.
NOTE: Press Drive to toggle the drive motor between Manual on, Manual off or Automatic mode.
Calibrating the Development Time
1.Run a cleanout sheet through the processor and use a stopwatch to determine the actual time taken from entering the developer tank to entering the bleach tank. If necessary, calibrate the processor.
2.Select the Drive screen.
3.Press Set twice to access the drive screen in the Calibrate mode.
4.Use the Up or Down Arrow to change the offset. As the offset changes, the displayed development time changes. Increase and decrease the offset until the displayed time matches the actual time. If necessary, choose an average value for the offset.
5.Record and save the calibration value.
6.Press Set twice or press a different blue function key to return to View mode.
October 2002 |