Maintaining the Equipment

Replacing the (Round) Balance Filters

IMPORTANT: The balance filters are made of glass and are expensive to replace. They are not included in the product warranty or service contract. Use care when handling the balance filters.

Wear white cotton gloves for this procedure.

It is not necessary to power down the scanner before changing the balance filters.

1.Open the lamp-and-filter-access door (see page 4-4for the location of the lamp-and-filter-access door).

3 screws

balance-filter

 

 

assembly

red balance-filter assembly handle

2.Loosen the thumbscrews on the lamp cover and pull the cover straight out.

When you remove the cover, the safety switch is released, removing power to the lamp socket.

CAUTION: Wait at least five minutes for the lamp to cool.

4-6