Maintaining the Equipment

3.Position the red balance-filter selection handle of the balance-filter assembly in the center.

4.Hold the red handle and pull the assembly straight out from the scanner.

5.Push the IR filter up from below the filter assembly.

6.Clean the IR filter with the camel’s-hair brush or, if needed, replace the IR filter as described in step 7.

CAUTION: Use care with the IR filter. It is not secured in place.

7.If replacing the IR filter, set the new filter in place with the arrows on the edges of the filter pointing down toward the scan lamp.

8.Install the filter assembly in the scanner with the balance-filter selection handle in the middle.

9.Reinstall the lamp cover, tighten the thumbscrews, and select a filter position.

10.Calibrate the scanner (see Calibrating the Scanner on page 3-4).

August 2001

4-9

Page 49
Image 49
Kodak HR 500 manual Maintaining the Equipment