CA LI BRATION AND TR OUBLESHOOTING

Calibration Overview

C H A P T E R 6

C H A P T E R 6

CALIBRATION AND TROUBLESHOOTING

This chapter describes calibration of your LDX-3232 Series Current Source. Descriptions of the required test instruments, calibration conditions, and the detailed procedures for calibration of the instrument's Laser Diode Current Source are included. A troubleshooting guide is also offered for some of the more common failure symptoms. This chapter is divided into three major sections, local (front panel) calibration, remote (through the GPIB Interface) calibration, and a troubleshooting guide.

Calibration Overview

There are several calibrations to be performed to completely calibrate your LDX-3232 Current Source. The Laser Diode Current Source calibration consists of calibrating the current source in both ranges and both modes (low and high bandwidth), calibrating the photodiode monitor feedback ammeter and finally the forward voltage measurement. The current limits are calibrated internally by the instrument as part of the calibration process.

Your LDX-3232 Current Source can be calibrated with the case closed. It should be calibrated every 12 months or whenever performance verification indicates that calibration is necessary, such as differences between setpoint and measurement display values that exceed the accuracy specification. Calibrate the instrument under laboratory conditions. We recommend calibration at 23°C ± 1.0°C. When necessary, however, the LDX-3232 Current Source may be calibrated at its intended use temperature if this is within the specified operating temperature range of 0 to 40°C.

04_06

LDX-3232

89

Page 105
Image 105
Lightwave Communications LDX-3232 manual Calibration and Troubleshooting, Calibration Overview