Lightwave Communications LDX-3232 manual LASerCALLDI?, LASerCALLDV, Ldi Las Cal, Las Cal Ldv

Models: LDX-3232

1 118
Download 118 pages 53.85 Kb
Page 60
Image 60

C H A P T E R 4

COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

LASer:CAL:LDI?

 

 

 

 

FRONT PANEL

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

REMOTE

 

 

 

The LASer:CAL:LDI? query is used to determine that the LDX-3232 is ready for a value

 

 

to be entered during the calibration cycle of the LASer:CAL:LDI mode.

 

Syntax Diagram

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LDI

 

 

:

LAS

 

 

:

CAL

:

 

I

?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Notes

After this query is issued and a response of 1 is received, the LDX-3232 will be ready for

 

 

the user to enter a current value via the LASer:LDI command (see Chapter 6).

 

Examples

"LASer:CAL:LDI?"

-response: 1, means the LDX-3232 is ready for the user to enter a

current value via the LASer:LDI command.

“LASer:Cal:LdI?” -response: 0, means the LDX-3232 is not yet ready for the user to enter a LASER current value

LASer:CAL:LDV

 

 

 

 

FRONT PANEL

 

 

 

 

 

 

 

 

 

 

 

 

REMOTE

 

The LASer:CAL:LDV command is used to enter the LASER voltage measurement

 

calibration mode.

 

 

 

 

 

Syntax Diagram

 

 

 

 

 

 

:

LAS

:

CAL

:

LDV

Parameters

None

 

 

 

 

 

Notes

After this command is issued, the LDX-3232 will allow calibration of the laser voltage

 

measurement. This procedure is outlined in Chapter 6.

 

 

 

The LASer:CAL:LDV? query may be used to determine if the LDX-3232 is ready for the

 

user to enter a value.

Examples

"Las:CAL:Ldv" -action: the LDX-3232 enters calibration mode for LASER voltage.

 

"Las:Cal:LDV" -action: the LDX-3232 enters calibration mode for LASER voltage.

44 LDX-3232

Page 60
Image 60
Lightwave Communications LDX-3232 manual LASerCALLDI?, LASerCALLDV, Ldi Las Cal, Las Cal Ldv