LDX-3232 Device-Dependent Command Reference

"LAS:Set:MDI?" -response: 100.0 means the laser monitor photodiode current is setpoint is 100 µA, for use in constant P mode with CALMD=0.

LASer:SET:MDP?

FRONT PANEL

 

 

REMOTE

The LASer:SET:MDP? query returns the laser monitor photodiode power setpoint value (when CALMD [CAL PD] is not zero), in mW.

:

LAS

:

SET

:

 

MDP

P

?

 

- where the response is the constand PPD setpoint, in mW.

Notes

In local operation, the MDP setpoint is read by selecting P mode, then pressing and

 

holding down the SELECT switch in the DISPLAY area of the front panel. When the mW

 

indicator is blinking, the display will show the MDP setpoint. This setpoint is used in

 

constand MDP mode only.

 

 

Examples

"LAS:Set:MDP?" -response: 10.0, means the laser monitor photodiode feedback

 

setpoint is 10.0 mW (CALPD>0).

 

 

 

"Laser:Set:MDP?" -response: 25.0, means the laser monitor photodiode feedback

 

setpoint is 25.0 mW (CALPD>0).

 

 

LASer:STEP

FRONT PANEL

 

 

REMOTE

The LASer:STEP command is used to increment or decrement the selected laser control mode setpoint by the given amount, when used with the LASer:INC or LASer:DEC command.

:

LAS

:

STEP

<white

space>

<nrf

value>

Parameters

An integer value of the step amount, in the range 1 to 9999.

Notes

The step of one corresponds to the smallest display resolution for the mode. For

 

example, a step of one means 0.1 mA, 0.1 mW, or 1 µA (if CALPD=0).

Examples

"Las:Mode:i; Las:ldi 20; Las:Step 100; Las:Inc; Las:set:ldi?" -action: sets the step to

 

10.0 mA, so the Las:set:ldi? query will return a value of 21.0 mA.

 

"LAS:STEP 100" -action: sets the step size to 100; could mean 10.0 mA , 10.0 mW, or

 

100 µA.