COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

C H A P T E R 4

LASer:MDLN

FRONT PANEL

 

 

REMOTE

The LASer:MDLN command enables or disables (turns off) the front panel modulation input (BNC).

Syntax Diagram

:

LAS

:

MDLN

<white

space>

<nrf

value>

Parameters

An <nrf value> - where 1=on, 0=off.

Notes

After turning the modulation off, it will remain off until turned back on via this command or

 

by resetting the instrument.

Examples

“las:mdln 1” -action: turns the modulation input on and enables modulation via the front

 

panel BNC connector.

 

“Laser:mdln 0” -action: turns the modulation input off. Modulation signals at the front

 

panel BNC will not effect the LAS output.

LASer:MDLN?

FRONT PANEL

 

 

REMOTE

The LASer:MDLN? query returns that status of the LAS modulation control. When disabled, a modulation signal at the front panel BNC will not effect the LAS output.

Syntax Diagram

 

:

LAS

:

MDLN

?

 

- where the response of 1=on, 0=off.

 

 

Notes

The LASER modulation control switches the front panel BNC in and out of the LAS

 

control circuit. At power up or after reset, modulation is enabled by default.

Examples

“LAS:MDLN?” -response: 1, means that the LAS modulation input is enabled.

 

“las:mdln?” -response: 0, means that the LAS modulation input is disabled.