REMOTE OPERATION

Error Messages

C H A P T E R 3

(up to 10 may be stored). Reading the error queue via GPIB clears the error queue.

Table 3.4 Error Code Classifications

Error Code Range

Area of Operation

E-001 to E-099

Internal Program Errors

 

 

E-100 to E-199

Parser Errors

 

 

E-200 to E-299

Execution Control Errors

 

 

E-300 to E-399

GPIB Errors

 

 

E-500 to E-599

LASER Control Errors

 

 

Table 3.5 LDX-3232 Series Error Codes

Error Code

Explanation

E-001

Memory allocation failure

 

 

E-103

<DEFINITE LENGTH ARBITRARY BLOCK PROGRAM DATA> length too long

 

 

E-104

<NON-DECIMAL NUMERIC PROGRAM DATA> type not defined

 

 

E-105

<DECIMAL PROGRAM DATA> exponent not valid

 

 

E-106

<DECIMAL PROGRAM DATA> digit expected

 

 

E-123

Command not found

 

 

E-124

<program mnemonic> Lookup, failed because query/command type match failed

 

 

E-126

Too few or too many program data elements

 

 

 

 

E-201

<PROGRAM DATA> value out of range

 

 

E-202

<PROGRAM DATA> will not convert to valid type

 

 

E-203

Security violation, command is not available without clearance

 

 

E-205

<PROGRAM DATA> is not a Boolean value or word

 

 

E-207

<PROGRAM DATA> will not convert to an unsigned 16-bit value

 

 

E-209

<PROGRAM DATA> will not convert to an unsigned 32-bit value

 

 

E-210

<PROGRAM DATA> will not convert to a floating point value

 

 

E-211

<PROGRAM DATA> will not convert to a character value

 

 

E-213

<PROGRAM DATA> is incorrect block data length

 

 

E-214

<PROGRAM DATA> length exceeds maximum

 

 

E-302

Query error; device was addressed to talk, but GPIB controller failed to read all of the

<RESPONSE MESSAGE>

04_06

LDX-3232

33

Page 49
Image 49
Lightwave Communications LDX-3232 manual Error Code Classifications, Response Message