COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

C H A P T E R 4

TERM?

 

 

 

 

 

 

FRONT PANEL

 

 

 

 

 

 

REMOTE

 

 

 

 

 

 

 

 

The TERM? query allows the programmer to determine whether the default

 

(IEEE488.2 standard) terminator has been altered to include a carriage return.

 

 

 

:

TERM

?

 

 

 

 

 

 

- where the response is 0 or 1.

 

 

 

Notes

An altered terminator will be in the form <CR><NL><^END>. This termination will be

 

sent with all output until the "TERM 0" command is sent, or the LDX-3232 is powered off.

Examples

"Term?"

-response: 1, means that the <CR><NL><^END> terminator is temporarily

 

being used.

 

 

 

 

"Term?" -response: 0, means that the <NL><^END> terminator (GPIB/IEEE488

 

standard) is being used.

 

 

 

 

 

 

 

 

 

 

 

TIME?

 

 

 

 

 

 

FRONT PANEL

 

 

 

 

 

 

REMOTE

 

 

 

 

 

 

 

 

The TIME? query allows the programmer to determine how much time has passed since

 

the LDX-3232 was last powered up.

 

 

:

TIME

?

 

 

 

 

 

 

 

- where the response is character data in the form: hours:minutes:seconds.

Notes

The TIME clock is independent of the TIMER clock.

Examples

"Time?" -response: 1:02.36, means that one minute and 2.36 seconds have passed

 

since the LDX-3232 was powered up.

"TIME?" -response: 32:00:76, means that 32 minutes and 0.76 seconds have passed since the LDX-3232 was powered up.