COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

C H A P T E R 4

LASer:LDV

FRONT PANEL

 

 

REMOTE

The LASer:LDV command sets the laser voltage for calibration of the laser voltage measurement.

Syntax Diagram

:

LAS

:

LDV

<white

space>

<nrf

value>

Parameters

An <nrf value> that represents the (laser) output voltage, in volts.

Notes

setpoint is the same for both low and high bandwidth output modes and is entered only

 

for LDV calibration.

Examples

"Las:LDV 4" -action: enters a value of 4.000 V.

 

"LAS:ldv 1.025" -action: enters a value of 1.025 V.

LASer:LDV?

FRONT PANEL

 

 

REMOTE

The LASer:LDV? query returns the value of the measured laser voltage. Syntax Diagram

 

:

LAS

:

LDV

?

 

- where the response is the measured laser output voltage for either low or high

 

bandwidth modes.

 

 

 

Notes

This measurement is updated approximately once every 600 ms.

Examples

"LAS:ldv?"

-response: 3.03, means the measured laser output voltage is 3.03 V.

"Laser:LDV?" -response: 1.0, means the measured laser output voltage is 1.000 V.

04_06

LDX-3232

63

Page 79
Image 79
Lightwave Communications LDX-3232 manual LASerLDV?, For LDV calibration, Las Ldv