C H A P T E R 4

COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

LASer:CAL:MDI?

FRONT PANEL

 

 

REMOTE

The LASer:CAL:MDI? query is used to determine that the LDX-3232 is ready for a value to be entered during the calibration cycle of the LASer:CAL:MDI mode.

Syntax Diagram

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MDI

 

:

LAS

 

:

CAL

:

 

 

?

 

 

 

 

 

 

 

 

 

 

 

 

 

IPD

Notes

After this query is issued and a response of 1 is received, the LDX-3232 be ready for the

 

 

user to enter a photodiode current value via the LASer:MDI command (see Chapter 6).

Examples

"LAS:CAL:MDI?" -response: 1, means the LDX-3232 is ready for the user to enter a

 

 

photodiode current value via the LASer:MDI command.

 

LASer:COND?

FRONT PANEL

 

 

REMOTE

The LASer:COND? query returns the value of the status condition register of the LASER operations.

Syntax Diagram

:

LAS

:

COND

?

-where the response is the sum of the following:

1

LASER Current Limit

256

Output is shorted

2

LASER Voltage Limit

512

Output is outside tolerance limit

4

N/A

1024

Output on/off state

8

Power limit

2048

Ready for calibration data state

16

Interlock disabled

4096

N/A

32

N/A

8192

N/A

64

Compliance Voltage Limit

16384

N/A

128

Open circuit

32768

N/A

Notes

The LASER conditions that are reported to the status byte are set via the

 

LASer:ENABle:COND command.

 

The Open circuit condition is only present while the LASER output is on, and when the hardware detects this condition, it will turn the LASER output off. Therefore, the Open

46 LDX-3232