Lightwave Communications manual LDX-3232 Current Source Command Set, Command Syntax

Models: LDX-3232

1 118
Download 118 pages 53.85 Kb
Page 38
Image 38

C H A P T E R 3

REMOTE OPERATI ON

LDX-3232 Current Source Command Set

Changing Operation from Local to Remote

Sending a command over the GPIB will automatically put the instrument in REMOTE mode. The REMOTE indicator will identify when the Current Source is in remote operation mode. When the instrument is in REMOTE mode, pressing the LOCAL switch returns the instrument to LOCAL control mode unless the Local Lockout state has been activated by the host computer. Local Lockout disables all LDX-3232 High Compliance Current Source front panel switches until this condition is changed by the host computer. When the instrument is put in Local Lockout Mode by the host computer, the REMOTE indicator will flash at a 1Hz rate to indicate that the front panel is completely disabled by Local Lockout.

The TALK/LISTEN indicator is illuminated when the instrument is communicating over the GPIB bus.

LDX-3232 Current Source Command Set

For the most efficient and effective remote control of your LDX-3232 High Compliance Current Source, we recommend you study the following sections. You will learn about the LDX-3232 command sets, both GPIB-IEEE488 Command Commands and the most used device specific commands, and command syntax. Later in this chapter you will be introduced to the error and status registers.

Command Syntax

Any command or command query must contain all of the letters that are shown in the LDX-3232 Command Path Structure, Figure 3.2. However, the Current Source's command parser is NOT CASE SENSITIVE so upper or lower case may be used in any combination. The lower case letters shown with the Device Dependent commands in Chapter 4 are optional letters and may be used for clarity. For example, the following commands are equal:

LAS:DIS 1

LAS:DisP 1

LAS:DISPLAY 1

In this example, only the first three letters "DIS" are required, while the other letters, "play" are optional. These optional letters must be used in the proper sequence as shown.

22 LDX-3232

Page 38
Image 38
Lightwave Communications LDX-3232 Current Source Command Set, Changing Operation from Local to Remote, Command Syntax