C H A P T E R 4

COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

LASer:RANge

FRONT PANEL

 

 

REMOTE

The LASer:RANge command selects the laser's drive current output range.

:

LAS

:

RAN

<white

 

 

 

 

space>

<nrf

value>

Parameters

An <nrf value> that represents the laser current output range.

 

2 = 2000 mA, and 4= 4000 mA range

Notes

This range setting effects the Laser Drive Current output current range only. The Laser

 

Drive Current output should be “off” when this command is issued. If the Laser Diode

 

Current output is “on” when this command is issued, the LDX-3232 will generate error

 

E515, and the range will not be changed.

Examples

"LAS:RAN 4" -action: sets the laser output drive current range to 4000 mA.

"Laser:range 2" -action: sets the laser output drive current range to 2000 mA.

LASer:RANge?

 

 

 

FRONT PANEL

 

 

 

 

 

 

 

 

 

 

REMOTE

 

The LASer:RANge? query returns the value of the Laser Drive Current range.

 

:

LAS

:

RAN

?

 

- where the response is in integer; 2 means 2000 mA range and 4 means 4000 mA

 

range.

 

 

 

 

Notes

The resolution and accuracy of the laser limit current are dependent on the Laser Drive

 

Current output range.

 

 

 

Examples

"LAS:RAN?"

-response: 2, means that the Laser Drive Current range is 2000mA.

"LAS:range?" -response: 4, means that the Laser Drive Current range is 4000mA.

74 LDX-3232

Page 90
Image 90
Lightwave Communications LDX-3232 manual LASerRANge?, Las Ran, Lasran?