C H A P T E R 6

CALIBRATION AND TR OU BLESHOOTING

Troubleshooting Guide

Symptom

Corrective Actions

Unable to adjust output

Check the I LIMIT parameter for the output range in use, see

 

that it is set above 0 mA.

Power mode operation has high

Check back panel PD BIAS; if set too low, may act as an open

output current, but little or no

feedback loop; if in doubt, set the PD BIAS to mid range (2.5 V)

power measured

 

 

 

Output exceeds power limit

The “power limit” is not a hardware limit. It only serves as a

 

warning that the power measurement has exceeded the limit

 

setpoint. Normally this limit will cause the output to shut off. This

 

feature can be defeated via the LAS:ENABLE:OUTOFF

 

command.

Open circuit error occurs during

Check load connections; check that measuring meter does not

calibration

auto-range (use non-auto-ranging modes).

 

 

Calibration is aborted

Calibration modes will be aborted if an open circuit is detected.

unintentionally

 

 

 

VOLTAGE LIMIT indicator blinks

This indicates a voltage limit error; check laser connections. A

 

high impedance may cause this condition.

Open circuit error E503 or voltage

The LDX-3232 instruments have an adjustable laser

limit error E505 prevents output

compliance voltage. Check to see if the LASER voltage limit

from reaching desired value

setting is too low (Chapter 2); check laser connections.

 

 

 

100LDX-3232

Page 116
Image 116