Lightwave Communications manual LDX-3232 Device-Dependent Commands, Command Description Format

Models: LDX-3232

1 118
Download 118 pages 53.85 Kb
Page 55
Image 55

COMMAND REFERENCE

LDX-3232 Device-Dependent Commands

LDX-3232 Device-Dependent Commands

C H A P T E R 4

This section contains all of the device-dependent commands for the LDX-3232 High Compliance Current Source, listed in alphabetical order. Sub-sections for each path are presented, listing the commands that are legal for that path. See Figure 3.1 for command path tree structure.

Command Path:Name Required letters are shown in upper case; lower case letters are optional.

Front Panel/

Remote Execution

A solid box means the command can be executed in that mode. Command mnemonics in front

Front Panel

Remote

LASer:LIMit:I1

panel (local) mode may vary.

Command Description

Syntax Diagram

(For Remote Commands)

Parameters

Tells what parameters are expected, and their type.

An <nrf value> refers to a flexible numeric notation (IEEE-488.2). Integer, exponential, or floating point may be used with the decimal radix. Alternate radices may be selected in remote operation.

Points of Interest

Has special information about the command.

Examples

Shows typical statements using the command.

The LASer:LIMit:I2 command sets the laser current limit value.

SYNTAX DIAGRAM

:

LAS

:

LIM

 

 

 

 

 

 

 

:

I2

<white

 

< nrf

 

 

 

space>

 

value >

 

 

 

 

 

 

 

PARAMETERS

An <nrf value> that represents the LASER limit

 

current, in mA.

 

 

 

 

POINTS OF

The current limit is in effect in all modes of

INTEREST

operation of the laser output.

 

In local operation, the limit current is entered

 

by selecting the LIM I parameter, pressing and

 

holding in the (PARAMETER) SET switch,

 

adjusting the ADJUST knob until the desired value

 

is displayed, and then releasing the SET switch.

EXAMPLES

"Las:Lim:I2 100" - action: the LASER current limit

 

is set to 100.0 mA.

 

 

 

"LAS:LIM:I4 3000" - action: the LASER current limit is set to 3000.0 mA.

Figure 4.1 Command Description Format

04_06

LDX-3232

39

Page 55
Image 55
Lightwave Communications manual LDX-3232 Device-Dependent Commands, Command Description Format