Lightwave Communications LDX-3232 LASerCALLDI, Selected. This procedure is outlined in Chapter

Models: LDX-3232

1 118
Download 118 pages 53.85 Kb
Page 59
Image 59

COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

C H A P T E R 4

LASer:CAL:LDI

FRONT PANEL

 

 

REMOTE

The LASer:CAL:LDI command is used to enter the LASER current setpoint, measurement, and limit (in low bandwidth mode) calibration mode.

Syntax Diagram

 

 

 

 

 

 

LDI

 

:

LAS

:

CAL

 

:

 

 

 

 

 

 

 

 

I

Parameters

None

 

 

 

 

 

Notes

Since the limit circuit is the same for both high and low bandwidth modes, it is only

 

calibrated when low bandwidth mode is selected.

 

 

 

After this command is issued, the LDX-3232 High Compliance Current Source will allow

 

calibration of the current setpoint, measurement, and limit (if low bandwidth mode is

 

selected). This procedure is outlined in Chapter 6.

 

 

 

The calibration defaults to the selected bandwidth setting. Therefore, it is necessary to

 

select the desired bandwidth (I or IHBW) and turn the LASER output on before performing

 

the calibration for that bandwidth.

 

 

 

 

 

Calibration is performed at the current setpoint, wherever it is set. If the LASER output is

 

not on, E516 is displayed.

 

 

 

 

 

In remote operation, the LASer:CAL:LDI? query may be used to determine if the

 

LDX-3232 High Compliance Current Source is ready for the user to enter a value.

Examples

"Las:CAL:LdI"

-action: the LDX-3232 enters calibration mode for LASER current.

"Las:Cal:LDI" -action: the LDX-3232 enters calibration mode for LASER current.

04_06

LDX-3232

43

Page 59
Image 59
Lightwave Communications LDX-3232 manual LASerCALLDI, Selected. This procedure is outlined in Chapter