C H A P T E R 6

CALIBRATION AND TR OU BLESHOOTING

Recommended Equipment

Finally, the LDX-3232 Current Source should be allowed to warm up for at least one hour before calibration.

Recommended Equipment

Recommended test equipment for calibrating the LDX-3232 Current Source is listed in Table 6.1. Equipment other than that shown in the table may be used if the specifications meet or exceed those listed.

Table 6.1 Recommended Test Equipment

Description

Manufacturer/Model

Specification

DMM

HP 3457A

DC Amps (@ 1.0 A): ±0.02%

 

 

Resistance (@ 10 Ω): ±0.02%

 

 

0.1 µA or 0.1 mV resolution

 

 

 

It will be necessary to connect various loads and circuits to the outputs of each current source for the calibration procedure. A schematic is shown in Figure 6.1 for the photodiode calibration circuit with the required components listed in Table 6.2. Also, the devices required for other calibration loads are listed in Table 6.2.

Table 6.2 Required Calibration Components

Laser Current Source Calibration

Current Source Calibration, all models

10

Ω, 20 W resistor, low TCR

 

 

 

Voltage Measurement Calibration

15

Ω, 5 W resistor, low TCR

 

 

 

PD Monitor Calibration (See Schematic in Figure 6.1)

 

 

 

 

PD Calibration Circuit

 

 

R1

 

49 Ω resistor, 1%, 1/4 W

R2

 

100 Ω resistor, 1%, 1/4 W

R4

 

1.0 MΩ resistor, 1%, 1/4 W

R3

 

5 Ω, 1%, 10W

U1

 

TIL 117 opto isolator

 

 

 

90 LDX-3232

Page 106