CA LI BRATION AND TR OUBLESHOOTING

Troubleshooting Guide

Troubleshooting Guide

C H A P T E R 6

This section is a guide to troubleshooting the LDX-3232 Current Sources. Some of the more common symptoms are listed here, and the appropriate troubleshooting actions are given. We recommend that the user start at the beginning of this guide. Read the symptom descriptions, and follow the steps for the corrective actions that apply. If you encounter problems that are beyond the scope of this guide, contact your ILX Lightwave customer service representative.

Symptom

Corrective Actions

GENERAL

 

 

 

LDX-3232 unit will not power up

Check AC power line voltage and power cord connection.

 

 

Power on, but display is frozen

This may occur if the instrument loses power (AC line) briefly.

and switches don’t operate

turn the power switch off and on again to restart.

 

 

Displays E514 on power up on

Hold (GPIB) LOCAL button, RANGE button and FINE button as

display

you are powering up the instrument. This clears the laser board

 

EEPROM, LLLLL will be displayed on the display; the laser

 

current source must be recalibrated.

Instrument resets at power up or

Hold (GPIB) LOCAL button, RANGE button and FINE button as

any LASER mode change

you are powering up the instrument. This clears the laser board

 

EEPROM, LLLLL will be displayed on the display; the laser

 

current source must be recalibrated.

LASER CURRENT SOURCE

 

 

 

Power on, but no current output

Check interlock pins on LASER input connector on instrument

 

rear panel; these pins must be shorted either directly or though

 

a switch.

 

If OPEN CIRCUIT indicator is lit, check the load connections

 

and then try again.

 

Check the OUTPUT ON switch, the corresponding LED should

 

be lit.

Output current at limit; can’t be

If POWER mode is used, check the monitor diode (feedback)

lowered

connections; try reversing the polarity of the monitor

 

photodiode; also check the photodiode bias adjustment on the

 

rear panel.

 

If in I or IHBW mode, check the current setpoint and I LIMIT

 

setting; setting the output below the limit may require several

 

turns of the adjust know if the setpoint is much greater than the

 

desired limit setting.

Output goes off intermittently

Check the interlock circuit; an intermittent interlock will turn the

 

output off.

 

Check that the AC power cord connection is secure; power line

 

drop-outs may reset the unit and when power is restored, the

 

output will be off.

04_06

LDX-3232

99

Page 115
Image 115