LASer:LIMit:I2

FRONT PANEL

 

 

REMOTE

The LASer:LIMit:I2 command sets the LASER current limit value for the 2000 mA range on the LDX-3232.

Syntax Diagram

:

LAS

:

LIM

:

I2

<white

space>

<nrf

value>

Parameters

An <nrf value> that represents the LASER limit current, in mA.

Notes

The current limit is in effect in all modes of operation (ILBW, IHBW, or P). The current

 

limit in use depends on the LASER output range selection.

Examples

"LAS:LIM:I2 180" -action: the LASER current limit is set to 180 mA.

 

":Laser:Limit:I2 160" -action: the LASER current limit is set to 160 mA.

LASer:LIMit:I2?

FRONT PANEL

 

 

REMOTE

The LASer:LIMit:I2? query returns the value of the LASER current limit for the 2000 mA range on the LDX-3232.

Syntax Diagram

 

 

 

 

 

 

 

 

 

 

:

LAS

 

:

LIM

:

 

I2

?

 

 

 

 

 

- where the response is the current limit value for the 2000 mA range.

 

Notes

The current limit is in effect in all modes of operation (ILBW, IHBW, or P). The current

 

limit in use depends on the LASER output range selection.

 

 

 

Examples

"LAS:LIM:I2?"

-response: 40, means the laser current limit is 40 mA.

 

"Laser:LIM:I2?" -response: 50, means the laser current limit is 50 mA .