Lightwave Communications LDX-3232 manual LASerCALMD?

Models: LDX-3232

1 118
Download 118 pages 53.85 Kb
Page 58
Image 58

C H A P T E R 4

COMMAND REFERENCE

LDX-3232 Device-Dependent Command Reference

LASer:CALMD?

 

 

FRONT PANEL

 

 

 

 

 

 

 

 

REMOTE

 

The LASer:CALMD? query returns the value of the laser's photodiode feedback

 

responsivity (CAL PD parameter) setting.

 

Syntax Diagram

 

 

 

 

 

 

 

CALMD

 

:

LAS

:

?

 

 

 

 

CALPD

Notes

If this value is 0, the LDX-3232 will be set to operate in constant IPD mode, and the IPD

 

setpoint value will be in effect. If this value is non-zero, the LDX-3232 will be set to

 

operate in constant PPD mode, and the PPD setpoint value will be in effect.

 

If this value is 0, the front panel PPD will display "-.-” and no PPD value can be calculated.

Examples

"LASER:CALMD?" -response: 0, means the LDX-3232 is set for operation in constant

 

IPD mode (if P mode is also selected).

 

"Las:Calmd?" -response: 1.1, means the LDX-3232 is set for operation in constant PPD

 

mode (if P mode is also selected) and the responsivity is set to 1.1 µA/mW. 1.1 µA of

 

photodiode feedback current represents 1 mW of optical power.

LAS:CAL:CONST?

FRONT PANEL

 

 

REMOTE

The LAS:CAL:CONST? query returns a stored calibration value. It is intended for factory use only.

42 LDX-3232

Page 58
Image 58
Lightwave Communications LDX-3232 manual LASerCALMD?, Responsivity CAL PD parameter setting Syntax Diagram