TRIGGER MENU | OPERATION |
NOTE
The averaging function averages a number of gate WIDTHS, so for a given averaging num- ber, larger WIDTHS will take longer to AVER- AGE. Narrower widths will average faster (but may yield a
NOTE
Use Arming to synchro- nize to other equipment or modulation/burst synchronization. This is a simple way to inhibit measurements during
DELAY | In Profile mode, DELAY sets the time delay (after the display trigger |
[TRGDLY | delay) to when the system starts to take and display readings, rep- |
GTDLY] | resented by the left most cursor. Enter 0.0 to 1.0 seconds, in ms or |
| s. |
| NOTE |
| Changing the left most cursor, or the trigger delay time, up- |
| dates the cursor or the delay time value accordingly. In Pro- |
| file mode, moving the cursor only allows updates to the |
| pixel resolution of the display. In Power vs. Time mode, the |
| delay and width can be used to alter the update rate. |
| In Readout mode (CUSTOM sensor mode), the value entered for |
| DELAY is applied after a trigger event, and before samples are |
| taken. Enter 0.0 to 1.0 seconds, in ms or s. |
WIDTH | Enter 100 ns to 7.0 seconds (the default is 20 ms). In Profile mode, |
[TRGGW | WIDTH is the gate time the system uses to perform a cursor aver- |
GTGW] | age measurement. The time interval is represented by the space |
| between the left most cursor and the right most cursor. Changing |
| either cursor, or the gate width value, updates both the cursors and |
| the gate width value. |
| In Readout mode, this value defines the measurement gate width. |
| A measurement is presented as the average of all data taken in this |
| gate width. |
| In Power vs. Time mode, the delay and width can be used to alter |
| the update rate or sample rate. |
EDGE | When set to External TTL, the power meter triggers on a TTL level |
[TRGXTTL | rising or falling. This selection sets the trigger for either a rising or |
GTXTTL] | falling edge. |
ARMING | Sets the trigger arming, unless the trigger source is set to EXTTTL. |
[TRGARM | When ARMING is set to Blanking ON, only samples taken when |
GTARM] | the rear panel Digital Input BNC is active will be averaged in the |
| measurement. The polarity of the rear panel Digital Input BNC sig- |
| nal can be set (high or low) using the SystemRear PanelBNCTTL |
| LEVEL menu setting. |
| When ARMING is set to Blanking OFF, all samples are read irre- |
| spective of the level on this BNC. |
| Figure |
ML2430A OM | |
|
|