TRIGGER MENU

OPERATION

NOTE

The averaging function averages a number of gate WIDTHS, so for a given averaging num- ber, larger WIDTHS will take longer to AVER- AGE. Narrower widths will average faster (but may yield a less-settled measurement).

NOTE

Use Arming to synchro- nize to other equipment or modulation/burst synchronization. This is a simple way to inhibit measurements during user-defined pe- riods without entering actual time periods.

DELAY

In Profile mode, DELAY sets the time delay (after the display trigger

[TRGDLY

delay) to when the system starts to take and display readings, rep-

GTDLY]

resented by the left most cursor. Enter 0.0 to 1.0 seconds, in ms or

 

￿s.

 

NOTE

 

Changing the left most cursor, or the trigger delay time, up-

 

dates the cursor or the delay time value accordingly. In Pro-

 

file mode, moving the cursor only allows updates to the

 

pixel resolution of the display. In Power vs. Time mode, the

 

delay and width can be used to alter the update rate.

 

In Readout mode (CUSTOM sensor mode), the value entered for

 

DELAY is applied after a trigger event, and before samples are

 

taken. Enter 0.0 to 1.0 seconds, in ms or ￿s.

WIDTH

Enter 100 ns to 7.0 seconds (the default is 20 ms). In Profile mode,

[TRGGW

WIDTH is the gate time the system uses to perform a cursor aver-

GTGW]

age measurement. The time interval is represented by the space

 

between the left most cursor and the right most cursor. Changing

 

either cursor, or the gate width value, updates both the cursors and

 

the gate width value.

 

In Readout mode, this value defines the measurement gate width.

 

A measurement is presented as the average of all data taken in this

 

gate width.

 

In Power vs. Time mode, the delay and width can be used to alter

 

the update rate or sample rate.

EDGE

When set to External TTL, the power meter triggers on a TTL level

[TRGXTTL

rising or falling. This selection sets the trigger for either a rising or

GTXTTL]

falling edge.

ARMING

Sets the trigger arming, unless the trigger source is set to EXTTTL.

[TRGARM

When ARMING is set to Blanking ON, only samples taken when

GTARM]

the rear panel Digital Input BNC is active will be averaged in the

 

measurement. The polarity of the rear panel Digital Input BNC sig-

 

nal can be set (high or low) using the SystemRear PanelBNCTTL

 

LEVEL menu setting.

 

When ARMING is set to Blanking OFF, all samples are read irre-

 

spective of the level on this BNC.

 

Figure 4-3 shows a typical arming diagram.

4-18

ML2430A OM

 

 

Page 46
Image 46
Anritsu ML2430A operation manual Shows a typical arming diagram