SOURCE SWEEP MODE | PROCEDURES |
| (a) CONNECT points. With this ON (default) the data points are connected with |
| vectors to resemble a real time trace. When OFF, the data points are displayed as |
| data points only, with no connecting line. This can give a faster display update, |
| however, it may be confusing as near vertical lines will have very few points de- |
| fined within them. |
| (b) If Tracking min/max is selected for the CHANNEL being used for the |
| PROFILE, it is possible to configure the tracking min/max to display the min and |
| max values for all the data BETWEEN THE CURSORS. This provides easy ac- |
| cess to peak values within a |
| When set to SINGLE it is updated EACH SWEEP and reflects the min and max |
| values only within that sweep. |
| When set to INFINITE, it maintains the min/max from the point it is started until it |
| is reset, updating the MAX if it sees a HIGHER measurement within the |
| CURSOR window, and updating the MIN readout if or when it sees a lower value |
| than that which it has already. This option, in the SYSTEM menu, is only a prefer- |
| ences option and not the main control for the feature. |
| The control for the min/max remains in the CHANNEL menu |
| |
| he wants to work and leave it. In most cases the SINGLE (default) is the most |
| useful as it provides a continuously updated readout of the min and max points |
| within the cursor window. The INFINITE setting is used when the results need to |
| be collated over a large number of samples. In order to RESET the INFINITE |
| configuration, use the CHANNEL menu. |
| NOTE |
| In the triggered modes, such as Internal A or B and Ex- |
| ternal TTL, the SENSORSETUPRange HOLD feature |
| can still be applied to restrict dynamic range if required. |
| This feature allows the ML2430A Series Power Meter to be synchronized to an |
MODE | RF source using the Horizontal ramp (to Analog Input) output on the RF source. |
| The ML2430A Series can also optionally use a Sequential Sync (to Digital Input) |
| output on the RF source. |
| To enable the Seq Sync input for RF sources that do provide this output (the de- |
| fault), set the SystemRear PanelBNCInput Port 1TTL LEVEL to HIGH. This |
| way the device will use both the Horizontal Ramp and Seq Sync inputs to sync to |
| the sweep. |
| To disable the Seq Sync input for RF sources that do not provide this output, set |
| the SystemRear PanelBNCInput Port 1TTL LEVEL to LOW. This way the de- |
| vice will only use the Horizontal Ramp to sync to the sweep. |
ML2430A OM | |
|
|