SOURCE SWEEP MODE

PROCEDURES

 

(a) CONNECT points. With this ON (default) the data points are connected with

 

vectors to resemble a real time trace. When OFF, the data points are displayed as

 

data points only, with no connecting line. This can give a faster display update,

 

however, it may be confusing as near vertical lines will have very few points de-

 

fined within them.

 

(b) If Tracking min/max is selected for the CHANNEL being used for the

 

PROFILE, it is possible to configure the tracking min/max to display the min and

 

max values for all the data BETWEEN THE CURSORS. This provides easy ac-

 

cess to peak values within a time-window; for example, the top of a pulse.

 

When set to SINGLE it is updated EACH SWEEP and reflects the min and max

 

values only within that sweep.

 

When set to INFINITE, it maintains the min/max from the point it is started until it

 

is reset, updating the MAX if it sees a HIGHER measurement within the

 

CURSOR window, and updating the MIN readout if or when it sees a lower value

 

than that which it has already. This option, in the SYSTEM menu, is only a prefer-

 

ences option and not the main control for the feature.

 

The control for the min/max remains in the CHANNEL menu

 

(CHANNELSETUP-more-MIN/MAX RESET). The user should select the way

 

he wants to work and leave it. In most cases the SINGLE (default) is the most

 

useful as it provides a continuously updated readout of the min and max points

 

within the cursor window. The INFINITE setting is used when the results need to

 

be collated over a large number of samples. In order to RESET the INFINITE

 

configuration, use the CHANNEL menu.

 

NOTE

 

In the triggered modes, such as Internal A or B and Ex-

 

ternal TTL, the SENSORSETUPRange HOLD feature

 

can still be applied to restrict dynamic range if required.

5-12SOURCE SWEEP

This feature allows the ML2430A Series Power Meter to be synchronized to an

MODE

RF source using the Horizontal ramp (to Analog Input) output on the RF source.

 

The ML2430A Series can also optionally use a Sequential Sync (to Digital Input)

 

output on the RF source.

 

To enable the Seq Sync input for RF sources that do provide this output (the de-

 

fault), set the SystemRear PanelBNCInput Port 1TTL LEVEL to HIGH. This

 

way the device will use both the Horizontal Ramp and Seq Sync inputs to sync to

 

the sweep.

 

To disable the Seq Sync input for RF sources that do not provide this output, set

 

the SystemRear PanelBNCInput Port 1TTL LEVEL to LOW. This way the de-

 

vice will only use the Horizontal Ramp to sync to the sweep.