GPIB OPERATION

HP 437B EMULATION COMMANDS

Logical OR

PowerOn

 

CommandError

ExecutionError

DeviceDependentError

 

 

 

7

6

5

4

3

2

1

0

7

6

5

4

3

2

1

0

Standard

Event Status Register

*ESR?Queue

not empty

Standard

Event Status Enable

Register

*ESE <NRf>

*ESE?

Output Queue

read by Serial Poll

Service

RQS

7 6 ESB MAV 3 2 1 0

Request

Generation

MSS

LogicalOR

 

 

 

 

 

 

7

5

4

3

2

1

0

Status Byte Register

read by *STB?

Service Request

Enable Register

*SRE <NRf>

*SRE?

Figure 6-8.IEEE 488.2 Standard Status Structures

*ESE?

Return Event status register enable mask

 

Syntax:

*ESE?

 

Remarks:

Returned format: <unsigned character>

 

 

When converted to an 8-bit binary number, this byte yields the bit settings

 

 

of the register.

*ESR?

Event status register request

 

Syntax:

*ESR?

 

Remarks:

Return the value of the standard event status register. Afterwards the event

 

 

status register are cleared. The returned format is: <unsigned character>.

When converted to a 8-bit binary number, this byte yields the bit settings of the register.

ML2430A OM

6-101

 

 

Page 188
Image 188
Anritsu ML2430A operation manual HP 437B Emulation Commands, Logical or