Anritsu ML2430A operation manual Related Commands FM, FH, FA, Related Commands TR

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 208
Image 208

GPIB OPERATION

HP 438A EMULATION COMMANDS

Remarks: Hold filter sets the filter mode to Manual from Auto, but retains the auto fil-

ter setting. This function is the same as the AVGM command.

Related

Commands: FM, FH, FA

FM

Set average value

Syntax: FM<val>EN

val: 0 to 9

Remarks: Sets the filter length for the averaging of sensor data. For HP 438A emula- tion, the filter length is defined as the number 2 to the power of <val>. For example, the command FM5EN would be 25, or 32.

Related

Commands: FH, FA

GT

Set group trigger

Syntax: GT<mode>

mode: 0 1 2

Remarks: 0 = Ignore Group Execute Trigger (GET) command 1 = Trigger immediate response to ‘GET’ command

2 = Trigger with delayed response to ‘GET’ command

The GTn command configures what the device does when it receives the ‘GET’ command.

For example: GT1 sets the’ GET’ (Group Execute Trigger) to perform a TR1 type trigger.

Related

Commands: TR

KB

Calibration factor

Syntax: KB<val><terminator>

val: 1.0 to 150.0% terminator: %

PCT

EN

ML2430A OM