GPIB OPERATION

ML4803A EMULATION COMMANDS

Set the offset value at the specified memory location in dBm.

Syntax: MOF<mem><;><value>

mem: Memory location 1 to 30.

value: Offset value in dBm.

Remarks: Set the offset value at memory store address <mem> to <value> dBm

MRF

Set the reference value at the specified memory location in

 

 

 

 

 

 

dBm.

 

 

 

 

 

 

 

 

 

 

 

Syntax:

MRF<mem><;><value>

 

 

 

 

 

 

 

 

mem:

Memory location 1 to 30.

 

 

 

 

 

 

 

 

value:

Reference value in dBm.

 

 

 

 

 

 

 

 

Remarks:

Set the reference value at memory store address <mem> to <value> dBm.

ODT

Output the current calibration factor, offset value, and refer-

 

 

 

 

 

 

ence level.

 

 

 

 

 

 

 

 

 

 

 

Remarks:

These are output as three separate messages in the output buffer, as

 

 

shown below:

 

 

 

 

 

 

 

 

 

 

CAL factor: 18 ASCII characters + <CR><LF>

 

 

 

 

 

 

 

OFFSET value: 18 ASCII characters + <CR><LF>

 

 

 

 

REFERENCE level: 19 ASCII characters + <CR><LF>

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1

5

10

 

15

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

C A L

 

 

 

.

 

d B

<CR><LF>

 

 

 

 

 

 

 

 

 

 

 

 

 

 

O F F S E T

 

 

.

 

d B

<CR><LF>

 

 

 

 

 

 

 

 

 

 

 

 

 

 

R E F E R E N C E

 

 

.

 

d B m <CR><LF>

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5 CHARACTER DATA

Figure 6-3.ODT Data Output Format

NOTE

When the ODT, OMR, and OPW data output commands are received simultaneously, only the command which is received last is valid.

ML2430A OM

6-89

 

 

Page 176
Image 176