Anritsu ML2430A operation manual Related Commands Llim, Mmrst, Mngdb, Syntax Mngd

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 140
Image 140

GPIB OPERATION

ML24XXA NATIVE COMMANDS

Remarks: The LLIMS command turns the limits on and off.

Related

Commands: LLIM

Query: LLIMS? <c>

Returned

String: LLIMS <c>,<state>

MMRST

Min Max Tracking reset

CHANNEL

Syntax: MMRST <c>

c:1 or 2

Remarks: This command resets the min/max values when in 'Readout' or 'Power vs.. Time' mode. In profile mode, this command is used to reset the channels min/max values.

MNGDB

Output Min Graph Binary Data

DATA OUTPUT

 

Syntax:

MNGDB

 

 

Remarks:

Available in graph modes only. Outputs in binary form the min graph data

 

 

to the GPIB in the long integer form of 1024 bits per dB as a definite length

arbitrary block response data. The C programming example 'Binary output decoding' on page 6-133shows how to extract the binary data. The re- sponse form is as follows :

MNGDB <#><length><number_of_bytes><data_byte_1><data_byte_2>

...<data_byte_n><\n>

<length> number of ASCII characters make up the number_of_bytes value <number_of_bytes> number of bytes of data contained in rest of the string <data_byte_n> four of these values makes up the long integer.

For example: FF FF D1 64 = –11932 As it is based on 1024 per dB, divide by 1024 to get the dB value (–11.652).

MNGD

Output Min Graph Data

DATA OUTPUT

Syntax: MNGD

Remarks: Available in graph modes only. Outputs in ASCII form the min graph data.

The format is as follows:

ML2430A OM

6-53

 

 

Page 140
Image 140
Anritsu ML2430A operation manual Related Commands Llim, Mmrst, Mngdb, Syntax Mngd