ML24XXA NATIVE COMMANDS

GPIB OPERATION

 

Syntax:

BAUTS <state>

 

 

state:

ENABLE or DISABLE

 

 

Remarks:

Enable/disable the battery auto power shut off.

 

 

 

NOTE: Although GPIB is not available under battery operation, the state of

 

 

this parameter can be changed for later use.

 

 

Query:

BAUTS?

 

 

Returned

 

 

 

String:

BAUTS <state>

 

BAUTT

Battery Auto shut off after x minutes

 

SYSTEM

 

Syntax:

BAUTT <val>

 

 

val:

10 to 240 minutes

 

 

Remarks:

Automatically turns the unit off after x minutes when operating on battery

 

 

power.

 

 

 

NOTE: Although GPIB is not available under battery operation, the state of

 

 

this parameter can be changed for later use.

 

 

Query:

BAUTT?

 

 

Returned

 

 

 

String:

BAUT <val>

 

BUFF

GPIB response buffering enabled

 

SYSTEM

 

Syntax:

BUFF <s>

 

 

s:

ON

 

 

 

OFF

 

 

Remarks:

If BUFF is ON: In the ML243X native mode, 488.2 GPIB operation, when a

 

 

request for data is made the response is put in an output buffer ready to be

read by the controller. If another data request is made and the previous data has not been read out of the output buffer; the new data is queued af- ter the original request. In this mode of operation the GPIB response buff- ering enable is ON, and following the 488.2 specifications when ever a request for data is made the response should be read.

If BUFF is OFF: In this mode when ever a request for data is made, (ex- cept by serial poll) the output buffer is cleared and the only data in the output queue will be the response to the last data request made. The out- put buffer is cleared once a valid GPIB data request command has been recognised.

6-20

ML2430A OM

 

 

Page 107
Image 107