Anritsu ML2430A operation manual Query OBZL?

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 147
Image 147

ML24XXA NATIVE COMMANDS

GPIB OPERATION

 

port:

1 or 2

 

 

val:

–5.00 to +5.00 Volts

 

 

Remarks:

Sets up the stop value for the voltage output in analog output mode. At-

 

 

tempting to set the start value to a voltage greater than the stop value, or

 

 

the stop value lower than the start value, will result in an execution error.

 

Query:

OBVSP? <port>

 

 

Returned

 

 

 

String:

OBVSP <port>,<val>

 

OBVST

BNC analog output start voltage scale

 

BNC

 

Syntax:

OBVST <port>,<val>

 

 

port:

1 or 2

 

 

val:

–5.00 to +5.00 Volts

 

 

Remarks:

Sets up the start value for the voltage output in analog output mode. At-

 

 

tempting to set the start value to a voltage greater than the stop value, or

 

 

the stop value lower than the start value, will result in an execution error.

 

Query:

OBVST? <port>

 

 

Returned

 

 

 

String:

OBVST <port>,<val>

 

OBZL

BNC RF blanking output level when zeroing

 

BNC

 

Syntax:

OBZL <level>

 

 

level:

HIGH (TTL high)

 

 

 

LOW (TTL low)

 

 

Remarks:

Sets the TTL level of the BNC RF blanking output.

 

Query: OBZL?

Returned

String: OBZL <level>

OFFCLR

Clear an offset table

SENSOR

Syntax: OFFCLR <val>

val: 1 to 5

Remarks: Sets all the values in the table specified to 0 dB and 0.00 Hz.

6-60

ML2430A OM

 

 

Page 147
Image 147
Anritsu ML2430A operation manual Query OBZL?