ML24XXA NATIVE COMMANDS

GPIB OPERATION

For example: FF FF D1 64 = –11932 As it is based on 1024 per dB, divide by 1024 to get the dB value (–11.652).

OGD

Output Graph Data

DATA OUTPUT

Syntax: OGD

Remarks: Outputs the next complete set of graph data. The format is as follows:

OGD <number_of_elements>,<element_1>,<element_2>,<ele- ment_n>...<\n>

The first number in the string is the number of elements to follow, and is al- ways 200 for the ML2430A Series.

OGSD

Output Valid Samples Array (power vs. time mode only)

DATA OUTPUT

Syntax: OGSD

Remarks: A power verses time chart plots the readings on a scrolling chart from left to right. If GRCP (connect points) is on and no new data has been re- ceived for a time slot, the graph is plotted with the same value as the previ- ous time slot but the data for that sample position is not marked as valid. This command will read out an array of 1's and 0's that indicate whether the data for that time slot is valid. For example, '1' for valid and '0' for con- necting data only.

Because the time between reading the data and reading the valid sample data may shift the valid samples out of sync with the graph read, it is rec- ommended that a 'HOLD ON' be issued before reading the graph and sample data, and a 'HOLD OFF' after. This will make sure that the sample data and the graph data agree.

If not in Power vs. Time mode, this command will set an execution error.

OI

Output device identification

 

GPIB 488.2

 

Syntax:

OI

 

 

 

Remarks:

Returned format:

 

 

 

 

<Company name>,<model>,<serial>,<firmware version>

 

 

ON

Output number of channel readings

 

DATA OUTPUT

6-64

ML2430A OM

 

 

Page 151
Image 151
Anritsu ML2430A operation manual Ogd, Syntax OGD, Syntax Ogsd, Syntax Remarks