GPIB OPERATION | ML24XXA NATIVE COMMANDS |
AVGLL | Auto low level averaging | SENSOR |
Syntax: AVGLL <s>,<mode>
s:A or B
mode: OFF LOW MEDIUM HIGH
Remarks: Sets the low level averaging window for the sensor. At resolution settings of 0.01 and 0.001dB, digital readouts may flicker due to the high reading rate of the ML2430A Series. Low level averaging applies a low pass filter to
± 0.01, 0.02, and 0.05 dB.
For example: When a LOW setting of low level averaging is applied while stepping from 0 dBm to
With a HIGH setting of low level averaging, the settling window is in- creased (up to approximately 0.05 dB) and the settling time is longer.
With low level averaging OFF, the meter displays the final reading instantly with no further settling observed. Any jitter due to noise is reflected in the displayed reading, which may be inconvenient for high resolution readings.
Example: AVGLL A,HIGH
Query: AVGLL? <S>
Returned
String: AVGLL <S>,<MODE>
AVGM | Manual Averaging | SENSOR |
Syntax: AVGM <s>
s:A or B
Remarks: Changes the averaging of the sensor to 'Moving' averaging mode from 'Auto' Averaging. The average number is set to the same value that the 'Auto' averaging mode was using internally. If the sensor is not presently in auto averaging mode, this command is ignored.
BAUTS | Battery Auto Turn OFF | SYSTEM |