OPERATION

SYSTEM MENU

CLEAR

[GPRST]

LINK CURSOR

[CURLK]

2

cursor 2 reading

￿P Power difference between cursor 1 and cursor 2

X1 X axis at cursor 1

X2 X axis at cursor 2

Data hold = MIN (or MAX) :

1cursor 1 minimum reading, or maximum if MAX mode

2cursor 2 minimum reading, or maximum if MAX mode

￿P Power diff between cursor 1 and cursor 2 minimums (or maximums if MAX mode)

X1 X axis at cursor 1

X2 X axis at cursor 2

Data hold = MIN&MAX:

1cursor 1 MIN reading cursor 1 MAX reading

2cursor 2 MIN reading cursor 2 MAX reading

Available in Profile, Source Sweep and Power vs. Time modes when the DATA HOLD representation selection is not set to NORMAL (or AVERAGE for Power vs. Time). Pressing the CLEAR key restarts the min/max collection.

Links the screen cursors in Profile and Power vs. Time modes so that when one is moved, both are moved. When the cursors are linked, a horizontal bar is drawn between them on the screen. If one cursor is moved, the other cursor moves with it to maintain their relative positions and time interval between them. When the cursors are linked, the relative time positions are altered by adjust- ing the gate width in the TRIGGERSETUPWIDTH submenu.

HOLD

[HOLD]

In Profile, Power vs. Time, and Source Sweep modes the graph HOLD function allows a graph to be held and printed. The key action is a toggle action, with the warning message Graph Display HELD displayed at the top of the screen when HOLD is active. Whenever measurement setup parameters are changed, graph hold will automatically be released.

AUTO scale Auto scale for all graphic modes (Profile, Source Sweep and Power [GRAUTO] vs. Time). In Profile and Source Sweep modes, auto scale will be

based on the min and max of the previous profile or sweep.

Display Controls the characteristics of the LCD display.

BACKLIGHT [DBLGHT]

Controls the LCD backlight during internal battery operation. Can be ON, OFF, or timed to go off after a specified period to save bat- tery life. The backlight is always on during AC or external DC power operation.

ML2430A OM

4-27

 

 

Page 55
Image 55
Anritsu ML2430A operation manual Clear