OPERATION

SENSOR MENU

Averaging

NOTE

Automatic averaging also applies an algo- rithm to enhance set- tling at low power levels (e.g.,signal sources).

SETUP

Sets up the Start and Stop frequencies and voltages when Source

[CVSTF

is set to V/GHz. This tells the ML2430A how to determine the fre-

CVSPF

quency of the swept signal based on the applied rear panel voltage.

CVSTV

 

CVSPV]

 

Sensor data averaging. The available soft keys depend upon the operating mode selected.

In Readout and Power vs. Time modes, the following soft keys appear:

SENSOR

Model ML2438A (dual channel) only. Select sensor A or B, in Power

 

vs. Time or Readout modes.

MODE

Select OFF, AUTO, MOVING or REPEAT, in Power vs. Time or

[AVG

Readout modes.

AVGM]

 

 

AUTOMATIC averaging uses a MOVING type of average and in-

 

creases the amount of averaging as the noise level increases. The

 

display updates at approximately 100 ms intervals, however the

 

data is available at the full rate. The display is slowed down to pre-

 

vent jitter and allow the user to follow the update.

 

MOVING average allows the user to manually select the amount of

 

averaging regardless of the signal level. The display is continually

 

updated while averaging.

 

When selected, the following soft key becomes available:

NUMBER

Sweep averaging number (1 to 512).

[AVG

 

AVGM]

 