GPIB OPERATION | ML24XXA NATIVE COMMANDS |
units: W (Watts) DBM (dB) DBUV (dBV) DBMV (dBmV)
Remarks: DBM 0dB is equal to 1mW readout mode W = Watts readout mode
V = Volts readout mode. This selection is automatically made when the channel input configuration is set to External volts (EXT V).
DBUV = dBV, 0dB is equal to 1V in readout mode.
Query: CHUNIT? <c>
Returned
String: CHUNIT <c>,<units>
When the channel input configuration is set to External volts (EXT V), the returned units are always volts, irrespective of what units have been set.
CONT | Continue | GPIB SETUP |
Syntax: CONT
Remarks: This command will allow the system to continue the startup sequence if there are self test failures other than DSP errors.
Related
Commands: STERR, START
CUR | Cursor in Power vs. Time and Source Sweep modes | SYSTEM | |
| Syntax: | CUR <cursor>,<fval> |
|
| cursor: | 1 or 2 |
|
| fval: | 0.0 to 1440 minutes (24 hours) in Power vs. Time mode |
|
|
| In Source Sweep mode, Power sweep : |
|
|
| In Source Sweep mode, Frequency sweep : 10.0 KHz to 122.0 GHz | |
| Remarks: | In Power vs. Time mode, the fval parameter is in minutes. In Source Sweep | |
|
| mode, the fval parameter is in dB or Hz for a power sweep or frequency | |
|
| sweep respectivily. |
|
Examples:
Power vs. Time:
Set cursor 1 to 30 seconds: CUR 1,0.5
Set cursor 2 to 12.5 hours: CUR 2,750
Source sweep:
Power Sweep, set cursor 1 to 11.5 dB: CUR 1,11.5
Frequency Sweep, set cursor 2 to 15.6 GHz: CUR 2,15.6GHz
ML2430A OM | |
|
|