GPIB OPERATION

ML24XXA NATIVE COMMANDS

units: W (Watts) DBM (dB) DBUV (dB￿V) DBMV (dBmV)

Remarks: DBM 0dB is equal to 1mW readout mode W = Watts readout mode

V = Volts readout mode. This selection is automatically made when the channel input configuration is set to External volts (EXT V).

DBUV = dB￿V, 0dB is equal to 1￿V in readout mode.

Query: CHUNIT? <c>

Returned

String: CHUNIT <c>,<units>

When the channel input configuration is set to External volts (EXT V), the returned units are always volts, irrespective of what units have been set.

CONT

Continue

GPIB SETUP

Syntax: CONT

Remarks: This command will allow the system to continue the startup sequence if there are self test failures other than DSP errors.

Related

Commands: STERR, START

CUR

Cursor in Power vs. Time and Source Sweep modes

SYSTEM

 

Syntax:

CUR <cursor>,<fval>

 

 

cursor:

1 or 2

 

 

fval:

0.0 to 1440 minutes (24 hours) in Power vs. Time mode

 

 

 

In Source Sweep mode, Power sweep : –120.0 dB to 30.0 dB

 

 

 

In Source Sweep mode, Frequency sweep : 10.0 KHz to 122.0 GHz

 

Remarks:

In Power vs. Time mode, the fval parameter is in minutes. In Source Sweep

 

 

mode, the fval parameter is in dB or Hz for a power sweep or frequency

 

 

sweep respectivily.

 

Examples:

Power vs. Time:

Set cursor 1 to 30 seconds: CUR 1,0.5

Set cursor 2 to 12.5 hours: CUR 2,750

Source sweep:

Power Sweep, set cursor 1 to 11.5 dB: CUR 1,11.5

Frequency Sweep, set cursor 2 to 15.6 GHz: CUR 2,15.6GHz

ML2430A OM

6-29

 

 