DE

display enable

Syntax: DE

Remarks: Return the display to normal operation after the display has been set in DD mode.

Related

Commands: DD, DF

DF

Display disable

Syntax: DF

Remarks: Turns the display off to allow faster measurements to be taken.

Related

Commands: DD, DE

DR

Set GPIB address

Syntax: DR<val><terminator>

val: 1 to 30

terminator: EN

Remarks: Changes the device address. The power meter default address is 13.

DY

Duty Cycle

Syntax: DY<val><terminator>

val: duty cycle value in percent terminator: %, PCT, or EN

Remarks: Sets the duty cycle to be applied to the input signal.

EMUL

GPIB emulation mode

Syntax: EMUL <mode>

mode: ML24XX (Anritsu ML2430A Series native mode) HP436A (Hewlett-Packard)

HP437B (Hewlett-Packard)