HP 438A EMULATION COMMANDS

GPIB OPERATION

Remarks: The calibration factor compensates for mismatch losses and effective effi- ciency over the frequency range of the power sensor.

Examples: KB99.9%

KB99.9PCT

KB99.9EN

LG

Set log units

Syntax: LG

Remarks: Changes the display to log units (dBm).

LH

Set high limit

Syntax: LH<val>EN

val: –99.999 to +99.999 (dBm only)

Remarks: Sets the high limit.

Example: LH30.00EN

LL

Set low limit

Syntax: LL<val>EN

val: –99.999 to +99.999 (dBm only)

Remarks: Sets the low limit.

Example: LL20.00EN

LM

limits check state

Syntax: LM<state>

state: 0 (off) or 1 (on)

Remarks: Turns limit checking on or off.

6-122

ML2430A OM

 

 

Page 209
Image 209
Anritsu ML2430A operation manual Examples KB99.9%, KB99.9PCT