Anritsu ML2430A operation manual Returned String Obdst port,units,val, BNC output mode select

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 146
Image 146

GPIB OPERATION

ML24XXA NATIVE COMMANDS

 

Query:

OBDST? <port>

 

 

Returned

 

 

 

String:

OBDST <port>,<units>,<val>

 

OBMD

BNC output mode select

 

BNC

 

Syntax:

OBMD <port>,<mode>

 

 

port:

1 or 2

 

 

mode:

'OFF' (output set to ground) port 1 or 2

 

 

 

'AOUT' (analog scaled output) port 1 or 2

 

 

 

'PASS/FAIL' (pass/fail) port 1 or 2

 

 

 

'SIGA' (signal output sensor A) port 1 only

 

 

 

‘LVLA1’ Signal channel range 1 amplifier output for sensor A

 

 

 

‘LVLA2’ Signal channel range 2 amplifier output for sensor A

 

 

 

‘LVLB1’ Signal channel range 1 amplifier output for sensor B

 

 

 

‘LVLB2’ Signal channel range 2 amplifier output for sensor B

 

 

 

'ACMOD' (AC mod output) port 1 only

 

 

 

'RFB' (RF blanking while zeroing) port 2 only

 

 

 

'SIGB' (signal output sensor B) port 2 only

 

 

Remarks:

Changes the type of output selected for the BNC outputs.

 

 

Query:

OBMD? <port>

 

 

Returned

 

 

 

String:

OBMD <port>,<mode>

 

OBPL

BNC pass/fail pass level

 

BNC

 

Syntax:

OBPL <port>,<level>

 

 

port:

1 or 2

 

 

level:

HIGH (TTL high is PASS)

 

 

 

LOW (TTL low is PASS)

 

 

Remarks:

Selects the PASS level for the Pass/fail type of output.

 

 

Query:

OBPL? <port>

 

 

Returned

 

 

 

String:

OBPL <port>,<level>

 

OBVSP

BNC analog output stop voltage scale

 

BNC

 

Syntax:

OBVSP <port>,<val>

 

ML2430A OM

6-59

 

 

Page 146
Image 146
Anritsu ML2430A Returned String Obdst port,units,val, BNC output mode select, Returned String Obmd port,mode