BATTERY CHARGING, REMOVAL AND REPLACEMENT

INSTALLATION

 

2-7BATTERY

The optional ML2430A Series Power Meter battery is a 12 Volt, 3000

 

CHARGING,

mA-h nickel-metal hydride (Ni-MH) multi-cell pack, located in a compart-

 

REMOVAL AND

ment on the bottom of the housing. The compartment cover is secured by

 

REPLACEMENT

fractional turn fasteners, as shown in Figure 2-1.Rotate the fasteners ap-

 

 

proximately ¼-turn counterclockwise to release the cover.

 

 

NOTE

 

 

 

 

 

 

 

 

 

 

ML2430A

 

BATTERY

BATTERY

 

 

 

COVER

 

 

 

OPEN

OPEN

BATTERY

COVER

FASTENERS

NOTE: Do not lubricate the battery cover fasteners.

Figure 2-1.Model ML2430A Series Battery Compartment

The battery is shipped with a partial charge only, and should be fully charged before use.

The battery can be completely charged in about two hours with the power meter in standby mode by selecting CHARGE from the System menu (page 4-30). This selection is available only when the instrument is being powered by AC line power or external DC power greater than 21 volts. Note that the instrument will shut down during the charging cycle, and re- start automatically when the charging is completed. A series of 10 beeps signals completion of the charge cycle.

The external battery charger (option ML2400A-13) can completely charge the battery in 2.5 hours.

ML2430A OM