Anritsu ML2430A operation manual Syntax RLmode, Mode 0 1, RM3EN, Related Commands RH, Syntax RV

Models: ML2430A

1 271
Download 271 pages 6.82 Kb
Page 200
Image 200

GPIB OPERATION

HP 437B EMULATION COMMANDS

RL

Relative mode

Syntax: RL<mode>

mode: 0 1 2

Remarks: Relative mode permits any measurement result to be compared in dB or percent to a reference value. Relative mode can be enabled using the cur- rent power reading (RL1) or the previous reference level (RL2). Successive measurements are displayed relative to this reference value. RL0 disables relative mode.

RM

Range hold set

Syntax: RM<val>EN

val: 0 to 5

Remarks: Set the range to <val> and then sets range hold. A value of 0 selects Auto Ranging, so that the range will change to take the best measurement auto- matically.

Example: To set the range to 3:

RM3EN

Related

Commands: RH

RV

Service request mask value.

Syntax: RV

Remarks: Read service request mask value. The returned string format is: <integer value>

Converting the integer value into an 8-bit binary number, each bit corre- sponds to the Service Request mask bits.

SE

Select cal factor table

Syntax: SE<table_number>EN

table number: 0 to 9

ML2430A OM