GPIB OPERATION

HP 436A EMULATION COMMANDS

6-13HP 436A

This section provides an alphabetical listing of the commands (mnemon-

EMULATION

ics) used to program the Model ML2430A Series Power Meter when in HP

COMMANDS

436A Emulation mode. The emulation mode is set through the front panel

 

SYSTEMRear PanelGPIBMODE menu (see Chapter 4, Operation) or

 

through the GPIB command EMUL (page 6-96).

 

HP Emulation commands must not have a space between the command

 

header and the parameter, or commas between the parameters.

 

The format for HP Emulation commands is:

 

<command header><parameter 1><parameter n>...

 

The end of the command text must be terminated with a line feed charac-

 

ter (0Ah, decimal 10) or a GPIB End of Transmission State (EOI), or both.

+Disable cal factors

Enable cal factors

1, 2, 3, Set sensor operating range 4 & 5

Remarks: Range 5 is the highest power range, range 1 the lowest. (These are the opposite to the ML2430A native mode ranges; that is, HP 436 range 5 sets to ML2430A range 1, and HP 436 range 4 to ML2430A range 2, etc.)

When the ML2430A is being used to emulate the HP 436, the ranges are reversed; that is, HP 436 range 1 (the lowest power range) is equivalent to the ML2430A range 5, and HP 436 range 5 (the highest power range) is equivalent to the ML2430A range 1. Refer to page 4-6for more information on sensor ranges.

9

Auto range

Remarks: Sets the ML2430A Series to automatically select the correct range for the measurement.

AWatt

Remarks: Set units to Watts. Turn relative mode off and do not allow relative.

B

ML2430A OM

6-95

 

 

Page 182
Image 182