GPIB OPERATION

ML24XXA NATIVE COMMANDS

Remarks: Normally when the system is powered on the ML2430A Series returns to the state it was in when it was powered off. This includes all the offset ta- bles, calibration adjust values, etc.

If Secure is set to ON, non-volatile memory is disabled and all stored val- ues are reset to the factory defaults when the system is powered on. As long as this command is ON, the system will load the presets (see Appen- dix A, Section A-3)every time it is turned ON.

Query: SECURE?

Returned

String: SECURE <s>

SENMM

Sensor Measurement mode

SENSOR

Syntax: SENMM <s>,<mode>

s:A or B

mode: DEFAULT (carrier wave) MOD (modulated average)

CUSTOM (user configurable trigger setup mode)

Remarks: Tells the sensor the type of signal it is expecting. This helps the sensor to take the best measurements.

Query: SENMM? <s>

Returned

String: SENMM <s>,<mode>

SENMO

Universal Sensor Operation Mode

SENSOR

Syntax: SENMO <s>,<value>

s:A or B

value: TRMS / FCW

Remarks: Selects between using a universal power sensor in its normal operating mode (TRMS) and its option 1 mode (FCW). FCW can only be selected if this option is fitted in the power sensor.

Query: SENMO? <s>

Returned

String: SENMO <s>,<value>

ML2430A OM

6-69

 

 