ML24XXA NATIVE COMMANDS | GPIB OPERATION |
After a TR2 command the instrument returns to either TR0 (trigger hold) or
TR3 (trigger free run) mode depending on what it was previously set to.
Related
Commands: TR0, TR1, TR3, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TR3 | Trigger free run | GPIB TRIGGER |
Syntax: TR3
Remarks: Sets the ML2430A Series back into free run mode on both channels.
Related
Commands: TR0, TR1, TR2, *TRG, Group Execute Trigger (GET), GT0, GT1, GT2
TRGARM Trigger arming | TRIGGER |
Syntax: TRGARM <c>, <state>
c:1, 2 or 1&2 state: ON or OFF
Remarks: Sets the readout trigger arming ON or OFF when in READOUT or POWER vs. TIME mode.
Select channel 1, 2 or 1&2. Selecting 1&2 allows both channels to trigger together on the same conditions without having to set up two sets of trig- ger data.
If set to ON, the system first checks to see if the BNC sweep blanking input is TRUE before it starts to trigger. If set to OFF, the system uses the trigger source (TRSRC) to decide when to trigger.
TRGARM will return an execution error if trying to set trigger arming ON when a display channel trigger source is already set to EXTTTL, as they both use the same BNC input.
Query: TRGARM? <c>
Returned
String: TRGARM <c>,<state>
The TRG type commands return the trigger state of the selected channel if the channel is ON. This depends on the settings of the “link triggers” flag and the current mode of the sensor on the selected channel. If the channel is OFF, the stored trigger state of the channel is returned.
TRGDLY | Trigger sample delay | TRIGGER |
Syntax: TRGDLY <c>, <val>[units]
ML2430A OM | |
|
|