ML24XXA NATIVE COMMANDS

GPIB OPERATION

state: ON or OFF

Remarks: This will link the trigger set-up between Profile mode and Readout mode so that the sample delay and the gate width will agree. A change to the trigger set-up in either Readout or Profile system set-up will affect either display mode.

Query: LINK?

Returned

String: LINK <state>

LLIM

Set Low limits

CHANNEL

Syntax: LLIM <c>,<val>

c:1 or 2

val:

Units

Min

Max

 

 

 

 

 

 

dBm

–99.99

+99.99

 

 

 

 

 

dBmV

–53.00

147.00

 

 

 

 

 

dBV

7.00

207.00

 

 

 

 

 

Watts

0.0

50.0

 

 

 

 

 

Volts

0.0

20.0

 

 

 

 

Remarks: Sets the low limit. The LLIMS command turns the limits on and off. It is not necessary to enter the units as the limit value is checked against the dis- played value.

Therefore, if the limits have been set for –10 dBm (LLIM 1, –10) and the display units are subsequently changed from dBm to Watts, the system still checks for the reading to rise above –10, even though the display units type has been changed.

Related

Commands: LLIMS

Query: LLIM? <c>

Returned

String: LLIM <c>,<val>

LLIMS

Turn on/off low limits

CHANNEL

Syntax: LLIMS <c>,<state>

c:1 or 2

state: ON or OFF

6-52

ML2430A OM

 

 

Page 139